Sign In | Join Free | My burrillandco.com
Home >

Nihon Kohden Cardiolife TEC-7621C Defibrillator Blood Pressure Plate UR-0257 6190-022986A S1

    Buy cheap Nihon Kohden Cardiolife TEC-7621C Defibrillator Blood Pressure Plate UR-0257 6190-022986A S1 from wholesalers
     
    Buy cheap Nihon Kohden Cardiolife TEC-7621C Defibrillator Blood Pressure Plate UR-0257 6190-022986A S1 from wholesalers
    • Buy cheap Nihon Kohden Cardiolife TEC-7621C Defibrillator Blood Pressure Plate UR-0257 6190-022986A S1 from wholesalers
    • Buy cheap Nihon Kohden Cardiolife TEC-7621C Defibrillator Blood Pressure Plate UR-0257 6190-022986A S1 from wholesalers
    • Buy cheap Nihon Kohden Cardiolife TEC-7621C Defibrillator Blood Pressure Plate UR-0257 6190-022986A S1 from wholesalers
    • Buy cheap Nihon Kohden Cardiolife TEC-7621C Defibrillator Blood Pressure Plate UR-0257 6190-022986A S1 from wholesalers

    Nihon Kohden Cardiolife TEC-7621C Defibrillator Blood Pressure Plate UR-0257 6190-022986A S1

    Ask Lasest Price
    Brand Name : Nihon Kohden
    Model Number : TEC-7621C
    Payment Terms : T/T, Western Union,PayPal
    Delivery Time : 3-5 working days
    MOQ : 1pcs
    • Product Details
    • Company Profile

    Nihon Kohden Cardiolife TEC-7621C Defibrillator Blood Pressure Plate UR-0257 6190-022986A S1


    Nihon Kohden Cardiolife TEC-7621C Defibrillator Blood Pressure Plate UR-0257 6190-022986A S1


    Brief Introduction

    This Blood Pressure Plate is work for Nihon Kohden Cardiolife TEC-7621C Defibrillator.

    CWT Technology HK Limited, located at Guangzhou, China. During these years of marketing and expanding, we have accumulated a great wealth of experience both in product distribution and customer service. We are a trust worthy medical equipment supplier as we have been engaged in this industry for so many years. Customer will received satisfactory goods.

    We understand the importance of the concept of win-win cooperation, we commit ourselves to protect the medical equipment market, satisfying all the customers' demands is our permanent pursuit.

    We are looking forward to establishing a long cooperative relationship with you.


    NIBP Specifications


    ParameterSpecification
    Display parametersSystolic pressure, mean pressure, diastolic pressure, cuff pressure
    Measuring range:0 to 300 mmHg
    Accuracy:±3 mmHg (0 mmHg ≤ NIBP < 200 mmHg)
    ±4 mmHg (200 mmHg ≤ NIBP ≤ 300 mmHg)
    Measurement mode:Manual / STAT (continuous) / Automatic (periodic)

    Alarm:

    Adult

    • Upper limit range: 15 to 260 mmHg in 5 mmHg steps, OFF
    • Lower limit range: OFF, 10 to 255 mmHg in 5 mmHg steps

    Neonate

    • Upper limit range: 10 to 125 mmHg in 5 mmHg steps, OFF
    • Lower limit range: OFF, 5 to 120 mmHg in 5 mmHg steps


    NIBP Mesurement Troubleshooting


    MessagePossibleCause Action
    NIBP air hose disconnectedThe air hose is disconnected from the NIBP socket.Connect the air hose properly. When NIBP monitoring is not necessary, press the SILENCE ALARM key to silence the alarm. The NIBP display can be cleared.
    The air hose is damaged.Replace the air hose with a new one.
    Faulty NIBP unit.Replace the NIBP unit.
    NIBP air leakThe cuff pressure does not change after inflation even after a certain period of time.Connect the cuff to the air hose properly.
    Connect the air hose to the socket properly.
    The cuff or air hose is damaged.Replace the cuff or air hose with a new one.
    Faulty NIBP unit.Replace the NIBP unit.
    NIBP measurement time outThe measuring time exceeded the specified time due to arrhythmia or noise.If the cause is arrhythmia, measure by invasive blood pressure measurement. Remove the cause if due to noise.
    Faulty NIBP unit.Replace the NIBP unit.
    Faulty NIBP unit.Replace the NIBP unit.

    Thank you for your time! If you are interested in Nihon Kohden Cardiolife TEC-7621C Defibrillator Blood Pressure Plate UR-0257 6190-022986A S1 , welcome to place your orders. Early order early shipment! If you want to explore more details please let us know.

    Quality Nihon Kohden Cardiolife TEC-7621C Defibrillator Blood Pressure Plate UR-0257 6190-022986A S1 for sale
    • Haven't found right suppliers
    • Our buyer assistants can help you find the most suitable, 100% reliable suppliers from China.
    • And this service is free of charge.
    • we have buyer assistants who speak English, French, Spanish......and we are ready to help you anytime!
    Submit Buying Request
    Send your message to this supplier
    *From:
    Your email address is incorrect!
    *Subject:
    Your subject must be between 10-255 characters!
    *Message:
    For the best results, we recommend including the following details:
    • --Self introduction
    • --Required specifications
    • --Inquire about price/MOQ
    Your message must be between 20-3,000
    Yes! I would like your verified suppliers matching service!
    Send your message to this supplier
     
    *From:
    *To: China World Technology Medical Equipment Service Group
    *Subject:
    *Message:
    Characters Remaining: (0/3000)
     
    Explore more Defibrillator Machine Parts products from this supplier
    Inquiry Cart 0