Sign In | Join Free | My burrillandco.com
Home >

Nihon Kohden Cardiolife TEC-7621C Defibrillator Relay Unit HV-761V With HV Drive Board UR-0121

    Buy cheap Nihon Kohden Cardiolife TEC-7621C Defibrillator Relay Unit HV-761V With HV Drive Board UR-0121 from wholesalers
     
    Buy cheap Nihon Kohden Cardiolife TEC-7621C Defibrillator Relay Unit HV-761V With HV Drive Board UR-0121 from wholesalers
    • Buy cheap Nihon Kohden Cardiolife TEC-7621C Defibrillator Relay Unit HV-761V With HV Drive Board UR-0121 from wholesalers
    • Buy cheap Nihon Kohden Cardiolife TEC-7621C Defibrillator Relay Unit HV-761V With HV Drive Board UR-0121 from wholesalers
    • Buy cheap Nihon Kohden Cardiolife TEC-7621C Defibrillator Relay Unit HV-761V With HV Drive Board UR-0121 from wholesalers
    • Buy cheap Nihon Kohden Cardiolife TEC-7621C Defibrillator Relay Unit HV-761V With HV Drive Board UR-0121 from wholesalers
    • Buy cheap Nihon Kohden Cardiolife TEC-7621C Defibrillator Relay Unit HV-761V With HV Drive Board UR-0121 from wholesalers

    Nihon Kohden Cardiolife TEC-7621C Defibrillator Relay Unit HV-761V With HV Drive Board UR-0121

    Ask Lasest Price
    Brand Name : Nihon Kohden
    Model Number : TEC-7621C
    Price : USD
    Payment Terms : T/T, Western Union, Paypal
    Delivery Time : 3-5 working days
    • Product Details
    • Company Profile

    Nihon Kohden Cardiolife TEC-7621C Defibrillator Relay Unit HV-761V With HV Drive Board UR-0121

    Relay unit HV-761V with HV Drive Board UR-0121

    Nihon Kohden Cardiolife TEC-7621C Defibrillator Relay Unit HV-761V with HV Drive Board UR-0121


    Brief Description

    Relay Unit HV-761V with HV Drive Board UR-0121 for Nihon Kohden Cardiolife TEC-7621C Defibrillator

    Relay Unit HV-761V with HV Drive Board UR-0121 - This part is second-hand but has complete performance and meets the usage requirements. It has been fully inspected by engineers before shipment to ensure its availability. You can always contact us before placing an order to know more details about this Relay Unit HV-761V with HV Drive Board UR-0121.


    ► Specification

    BrandNihon Kohden
    Product CategoryDefibrillation
    Product TypeSpare Parts
    Product NameRelay Unit with HV Drive Board
    Relay Unit P/NHV-761V
    HV Drive Board P/NUR-0121
    Machine ApplicationTEC-7621C
    ConditionUsed with Good Condition
    Warranty90 Days
    Delivery Date3 - 5 Working Days
    PackageCarton
    Payment MethodWestern Union, Bank T/T, Paypal
    TranspotationDHL, FedEx, UPS, TNT, EMS, By Air, etc.


    Romoving the HV Capacitor and Relay Unit

    1. Remove the battery pack, memory card and lower case.

    2. Remove the AC/DC unit.

    3. Remove the main chassis and motherboard.

    4. Short-circuit the two terminals of the high-voltage capacitor with a short bar, which is connected to the ground terminal on the wall through a ground wire.

    5. Remove the four M3 flat-head screws with spring washers that secure the high-voltage capacitor to the lower case, and then remove the two high-voltage capacitor retaining brackets.

    6. Remove the caps from the CNP552 and CNP553 wires from the high voltage capacitor.

    7. Remove the high voltage capacitor.

    8. Remove the three wires (CNP555, CNP554 and CNP556 wires) from the high voltage inductor that connect the high voltage inductor to the relay unit.

    9. Pull out the relay unit.

    NOTE: High voltage capacitors can store lethal energy. Always discharge the capacitor before touching any high-voltage component of the instrument (high-voltage capacitor, high-voltage inductor, relay unit, paddle socket, or paddle).

    More Details

    During the use of Nihon Kohden Cardiolife TEC-7621C Defibrillator, if there is a problem with the Relay Unit HV-761V, an error code will be displayed. Next, take a look at what the different error codes mean:

    1. A501: During standby mode, the HV capacitor has more than 1 J energy more than one continuous second

    2. A512: When charging starts, the high-voltage capacitor energy does not reach 1J within 2 seconds.

    3. A513: The energy did not reach the selected energy within the specified time.

    4. A524: After charging, the capacitor energy drops to the specified value for each energy.

    5. A527: After charging, the capacitor energy is approximately 15% higher than the selected energy.

    6. A529: After charging, the actual charging capacity is different from the selected capacity.

    7. A556: Internal discharge takes more than 20 seconds to complete.

    8. A566: 20 seconds after adjusting the internal discharge, the voltage of the high-voltage capacitor did not reach the target value.

    9. A585: The high voltage capacitor voltage exceeds the specified voltage.

    10. A587: When using disposable pads, 12.5% or more of the charged energy remains in the high-voltage capacitor after 2 seconds of external discharge.

    Detailed Photos


    Thanks for your interest in Nihon Kohden Cardiolife TEC-7621C Defibrillator Relay Unit HV-761V with HV Drive Board UR-0121, please feel free to contact me. Thank you for your time and wish you have a happy life.

    Quality Nihon Kohden Cardiolife TEC-7621C Defibrillator Relay Unit HV-761V With HV Drive Board UR-0121 for sale
    • Haven't found right suppliers
    • Our buyer assistants can help you find the most suitable, 100% reliable suppliers from China.
    • And this service is free of charge.
    • we have buyer assistants who speak English, French, Spanish......and we are ready to help you anytime!
    Submit Buying Request
    Send your message to this supplier
    *From:
    Your email address is incorrect!
    *Subject:
    Your subject must be between 10-255 characters!
    *Message:
    For the best results, we recommend including the following details:
    • --Self introduction
    • --Required specifications
    • --Inquire about price/MOQ
    Your message must be between 20-3,000
    Yes! I would like your verified suppliers matching service!
    Send your message to this supplier
     
    *From:
    *To: China World Technology Medical Equipment Service Group
    *Subject:
    *Message:
    Characters Remaining: (0/3000)
     
    Explore more Defibrillator Machine Parts products from this supplier
    Inquiry Cart 0