Sign In | Join Free | My burrillandco.com
Home >

Med-tronic Physio Control LIFEPAK 20 LP20 Therapy And Pacing Board 3202259 100-002-421F BMP006722-0067 21330-001560

    Buy cheap Med-tronic Physio Control LIFEPAK 20 LP20 Therapy And Pacing Board 3202259 100-002-421F BMP006722-0067 21330-001560 from wholesalers
     
    Buy cheap Med-tronic Physio Control LIFEPAK 20 LP20 Therapy And Pacing Board 3202259 100-002-421F BMP006722-0067 21330-001560 from wholesalers
    • Buy cheap Med-tronic Physio Control LIFEPAK 20 LP20 Therapy And Pacing Board 3202259 100-002-421F BMP006722-0067 21330-001560 from wholesalers
    • Buy cheap Med-tronic Physio Control LIFEPAK 20 LP20 Therapy And Pacing Board 3202259 100-002-421F BMP006722-0067 21330-001560 from wholesalers
    • Buy cheap Med-tronic Physio Control LIFEPAK 20 LP20 Therapy And Pacing Board 3202259 100-002-421F BMP006722-0067 21330-001560 from wholesalers
    • Buy cheap Med-tronic Physio Control LIFEPAK 20 LP20 Therapy And Pacing Board 3202259 100-002-421F BMP006722-0067 21330-001560 from wholesalers

    Med-tronic Physio Control LIFEPAK 20 LP20 Therapy And Pacing Board 3202259 100-002-421F BMP006722-0067 21330-001560

    Ask Lasest Price
    Brand Name : Med-tronic
    Model Number : Lifepak LP20
    Payment Terms : T/T, Western Union,PayPal
    Delivery Time : 3-5 working days
    MOQ : 1pcs
    • Product Details
    • Company Profile

    Med-tronic Physio Control LIFEPAK 20 LP20 Therapy And Pacing Board 3202259 100-002-421F BMP006722-0067 21330-001560


    Med-tronic Physio-Control LIFEPAK 20 LP20 Therapy Board and Pacing board 3202259 100-002-421F BMP006722-0067 21330-001560


    ► Brief Introduction

    The Therapy PCB controls the pacing and defibrillation therapy features. The primary communication between the Therapy PCB and the remainder of the device is through the data bus. A microprocessor and digital memory make up the central processing unit (CPU) that manages communication with the System Control PCB.


    Major subsystems

    Power Supplies — The Therapy PCB uses SW_VBatt (switched battery voltage) from the Power Module to originate five power supply voltages for use throughout the PCB as follows:

    • +5 V logic power to drive the CPU microprocessor and memory
    • ±15 V analog power for the pacing and therapy drive circuit
    • Patient-isolated 5 V analog power for the pacing and therapy circuits
    • Patient-isolated 15 V analog power for the pacing and therapy circuits
    • Patient-isolated 30 V analog power for the pacing and therapy circuits

    Power Switch — A power switch is a control circuit that detects the ON button selection from the User Interface PCB or a timer event from the System Control PCB to power up the device. This portion of the Therapy PCB is powered at all times, with very low quiescent current draw. When a power-on request is detected, this circuit switches VBatt (battery and/or ac converted dc power) provided by thePower Module to the remaining PCBs in the device. Low Battery (Battery Fail) is detected and a discrete signal is broadcast to other PCBs if battery voltage falls rapidly or reaches the point where normal operation is no longer feasible.


    Cap Charger — The cap charger is a high-voltage, patient-isolated circuit that charges the Energy Capacitor to the correct voltage for biphasic defibrillation (2 to 360 joules). Control is provided by the CPU, and capacitor voltage is provided back to the CPU for feedback. The cap charger is designed to nominally provide maximum charge rates and to automatically scale back to slower charge rates when low battery voltage is detected.


    Pacer Power Supply — The pacer power supply is a patient-isolated circuit that charges the Energy Capacitor up to the correct voltage for pacing. Control is provided by the CPU, and voltage regulation is maintained locally within the pacer supply. Capacitor voltage is provided back to the CPU for control through the cap charger circuitry


    H-Bridge — The H-Bridge is a patient-isolated circuit that creates the biphasic defibrillation waveform. A combination of silicon controlled rectifiers (SCR) and insulated gate bipolar transistors (IGBT) are used to place a positive-oriented defibrillation pulse across the patient load, followed immediately by a negative-oriented defibrillation pulse. The defibrillation pulse is delivered through the relay and Therapy Connector assembly to the external therapy cable on the outside of the device.


    Pacer — The pacer is a patient-isolated circuit that creates the pacing waveform. A portion of the H-Bridge circuitry is used to support the pacer by providing energy from the Defibrillation Capacitor. A current drive is used to control the amount of current provided to the patient during pacing.


    Relay — The relay provides patient isolation from the pacing and defibrillation circuitry when not in use. The relay is closed when the pacing current is set above zero and stays closed until the pacing current is set back to zero.


    Specification

    TypeLithium-ion, rechargeable
    VoltageTypical: 11.1 V
    Operating range: 9.0 V – 12.6 V
    Capacity2.4 Ah
    Operating Temperature0° to 45°C (32° to 113°F)
    Storage Temperature-20° to 60°C (-4° to 140°F) for up to 1 month
    Relative Humidity5 to 95% non-condensing


    Thank you for your time and interest in Med-tronic Physio-Control LIFEPAK 20 LP20 Therapy Board and Pacing board 3202259 100-002-421F BMP006722-0067 21330-001560. If you want to explore more details please let us know.


    Quality Med-tronic Physio Control LIFEPAK 20 LP20 Therapy And Pacing Board 3202259 100-002-421F BMP006722-0067 21330-001560 for sale
    • Haven't found right suppliers
    • Our buyer assistants can help you find the most suitable, 100% reliable suppliers from China.
    • And this service is free of charge.
    • we have buyer assistants who speak English, French, Spanish......and we are ready to help you anytime!
    Submit Buying Request
    Send your message to this supplier
    *From:
    Your email address is incorrect!
    *Subject:
    Your subject must be between 10-255 characters!
    *Message:
    For the best results, we recommend including the following details:
    • --Self introduction
    • --Required specifications
    • --Inquire about price/MOQ
    Your message must be between 20-3,000
    Yes! I would like your verified suppliers matching service!
    Send your message to this supplier
     
    *From:
    *To: China World Technology Medical Equipment Service Group
    *Subject:
    *Message:
    Characters Remaining: (0/3000)
     
    Explore more Defibrillator Machine Parts products from this supplier
    Inquiry Cart 0