Anesthetic Workstation X40, 8.4” LCD Human-computer interaction system with touch screen
anesthetic workstation X40
CO2 Absorber System
134°autoclavable CO2 absorber integrated bypass system, totally latex free.
Variable orifice flow sensor design.
35° (+-2) integrated heating system.
Bypass design with canister removes alarm.
No tools installation, easy maintenance.
System Specifications
| System Specifications | |
| Technical parameters | Specification | 
| Machine | |
| Size | 1410 × 950 × 650 (H × W × D) | 
| Weight | 110kg | 
| Top cover | 30kg bearing weight | 
| Size | 580 x 400mm | 
| Display | |
| Type | Color TFT (touch screen) | 
| Size | 8.4’ inches | 
| Resolution | 640 × 480 pixels | 
| Work surface light | LED | 
| LED indication | |
| Alarm Indication | |
| AC power | |
| Battery | |
| Audio instruction | |
| Speaker | Alarm sounds | 
| Control | |
| Button Press key | |
| Interface | |
RS232, VGA, USB Three auxiliary output power interface  | |
| Moving means | |
| Castor | 4 antistatic castors, diameter Φ125mm (2 brake Infront)  | 
| Toolbox | |
| Drawer | 200 × 392 × 398 mm(H×W×D)x2 | 
| Respiratory System | |
| Air bellow | 1500mL | 
| Absorption chamber | 1500mL | 
| Connection | Suction / ACGO ports: standard OD 22mm, ID 15mm, tapered connector; Exhalation ports: standard OD 22mm, ID 15mm, tapered connector. Manual breathing port: diameterΦ22cm  | 
| System leaks | < 40ml/min | 
| System compliance | ≤ 4 mL / cmH2O | 
| Resistance | Inspiratory < 0.6 kPa; Expiratory < 0.6 kPa  | 
| Vaporizer | |
| Description | Parameter | 
| Technical Specification | |
| Flowrate range | 0.2-15L/min | 
| Connector type | Selectatec compatible, plug in, cage mount | 
| Dosing methods | Pour-fil, Easy-fil, Quick-fil (Sevoflurane) | 
| Working environment | |
| Working temperature | +15℃ ~ +35℃ | 
| Relative humidity | 5%~95%(non-condensing) | 
| Atmospheric pressure | 70kPa ~ 106kPa | 
| Storage temperature | -20℃ ~ +55℃ | 
| Concentration: | 0~5.0%: Isoflurane, Halothane 0 ~ 8.0%: Sevoflurane  | 
| Main technical parameters | |
| Description | Parameter | 
| Display | 8” TFT, Touch screen, Top light | 
| Ventilation Model | VCV, PCV, Manual, Standby SIMV-V (Optional:SIMV-P, SPONT/PSV) | 
| Tidal volume | 20-1500ml(optional 10-1600ml), Electronic PEEP, Lung mechanics parameters. | 
| Wave form | Paw-T, Flow-T, V-T, CO2-T | 
| Loop | P-V, V-F | 
Pressure Monitoring  | Central: O2, N2O, Air Cylinder: O2, N2O  | 
Monitoring Parameter  | Tidal volume, MV, Frequency, I : E, Airway Pressure, Pmin, Compliance, Resistance | 
| Application | Adult and Infant | 
| Vaporizer | Selected bar for two position | 
| Gas supply | O2, N2O, Air | 
6 tubes Flowmeter  | O2: 0~10LPM; N2O: 0~12LPM; Air: 0~15LPM;  | 
| O2% | O2 sensor | 
| Auxiliary O2 | Auxiliary O2 Flow meter(O2: 0~15LPM) (Optional) | 
| ACGO | Yes | 
| Battery | NI-MH battery build-in. >90mins  | 
| Yoke | 2 ( O2, N2O) (Optional) | 
| Gas Cylinder | O2, N2O (Optional) | 
| Cycle Absorber | Intergrade, bypass Design, heating system. | 
| Setting Parameter | |
| Description | Parameter | 
| Tidal volume | 20~1500 mL(optional 10-1600ml) | 
| Frequency | 1~100 bpm | 
| Tinsp | 0.1~10.0s | 
| I:E | 4:1~1:10 | 
| Pause | 0~60% | 
| PEEP | OFF, 3~30 cmH2O | 
| Psupp | 0~70 cmH2O | 
| Pressure Control | 5~70 cmH2O | 
| Flow Trigger | 0.5~20 L/min | 
| Pressure Trigger | 0~20 cmH2O | 
| Ramp | 0~2s | 
| Monitoring Parameter | |
| Description | Parameter | 
| Tidal volume Inspiration | 0~2500 mL | 
| Tidal volume Expiration | 0~2500 mL | 
| MV | 0~60 L/min | 
| MVspont | 0~60 L/min | 
| Frequency | 0~100 bpm | 
| Ratespont | 0~100 bpm | 
| I : E | 9:1~1:99 | 
| Ppeak | 0~100 cmH2O | 
| Pmean | 0~100 cmH2O | 
| PEEP | 0~100 cmH2O | 
| Pplat | 0~100 cmH2O | 
| O2% | 15~100% | 
| Compliance | 0~300 mL/cmH2O | 
| Resistance | 0~600 cmH2O/(L/s) | 
| EtCO2 (Optional) | 0~13.3 % | 
| FiCO2 (Optional) | 0~13.3 % | 
| Alarm Parameter | |
| Description | Parameter | 
| Tidal volume | |
| Upper limit | 30~2000 mL | 
| Lower limit | OFF, 20~1500 mL | 
| MV | |
| Upper limit | 1~99 L | 
| Lower limit | 0~98 L | 
| O2% (Optional) | |
| Upper limit | 22~100%, OFF | 
| Lower limit | 20~99% | 
| Airway Pressure | |
| Upper limit | 10~99 cmH2O | 
| Lower limit | 1~98 cmH2O | 
| Frequency | |
| Upper limit | 1~100 bpm | 
| Lower limit | 0~99 bpm | 
| EtCO2 (Optional) | |
| Upper limit | 0.1~13.3 % ; | 
| Lower limit | 0~13.2 %; | 
| FiCO2 (Optional) | |
| Upper limit | 0.1~13.3 % ; | 
| Continue Pressure High | Airway pressure beyond(PEEP+15)cm H2O, continuedly(15+1)s | 
| Apnea | 10~60 s, Increase: 1s | 
| O2 supply down | < 0.28 MPa | 
| Mains failure | Automatic switch battery | 
| Battery low | < 10 min | 
| Battery discharged | < 5 min | 
| Mute | ≤120 s | 
| Pressure low | < -10cmH2O | 
| Power Specifications | |
| Description | Parameter | 
| External AC power | |
| Input Voltage | 100 - 240V | 
| Input Frequency | 50/60Hz | 
| Input Power | <150 VA | 
| Internal Battery | |
| Number of batteries | A battery pack | 
| Battery Type | Nimh batteries | 
| Rated battery voltage | 12VDC | 
| Battery capacity | 4200mAh | 
| Shutdown Delay | Less than 10min (using a new fully charged battery, low battery alert since the first post) | 
| Shortest supply time | 90min | 
| Charging time | 4h | 
| Work environment | |
| Temperature | 10 ~ 40 ºC | 
| Humidity | 5 ~ 95%, non-condensing | 
| Environmental pressures | 70 ~ 106 kPa | 
| Storage environment | |
| Temperature | -20 ~ 55 ºC | 
| Humidity | 5~ 95%, non-condensing | 
| Environmental pressures | 50 ~ 106 kPa | 

                                 
                             | 
                        