Display Blurry Screen ultrasonic repairing system Siemens S2000 Medical Solutions Replace RM Board
|
Display Blurry Screen Siemens S2000 Medical Solutions Replace RM Board 1. Repair time: 2-4 days 2. Warranty: 60 days 3. The video signal is output from RM to VI and then transmitted from VI to IO. You can start to check from the source RM. From RM to an external monitor, it is also found that the screen is blurred, so it can be determined that it is caused by a fault in RM. Replace the graphics card in RM, and the fault is eliminated. Repair...