Sign In | Join Free | My burrillandco.com
Home >

Drager Neonatal Straight Flow Sensor ISO15 8411130 For Straight Flow Sensor Babylog 8000 Evita4 Evita2 Dura NeoFlow

    Buy cheap Drager Neonatal Straight Flow Sensor ISO15 8411130 For Straight Flow Sensor Babylog 8000 Evita4 Evita2 Dura NeoFlow from wholesalers
     
    Buy cheap Drager Neonatal Straight Flow Sensor ISO15 8411130 For Straight Flow Sensor Babylog 8000 Evita4 Evita2 Dura NeoFlow from wholesalers
    • Buy cheap Drager Neonatal Straight Flow Sensor ISO15 8411130 For Straight Flow Sensor Babylog 8000 Evita4 Evita2 Dura NeoFlow from wholesalers
    • Buy cheap Drager Neonatal Straight Flow Sensor ISO15 8411130 For Straight Flow Sensor Babylog 8000 Evita4 Evita2 Dura NeoFlow from wholesalers
    • Buy cheap Drager Neonatal Straight Flow Sensor ISO15 8411130 For Straight Flow Sensor Babylog 8000 Evita4 Evita2 Dura NeoFlow from wholesalers
    • Buy cheap Drager Neonatal Straight Flow Sensor ISO15 8411130 For Straight Flow Sensor Babylog 8000 Evita4 Evita2 Dura NeoFlow from wholesalers

    Drager Neonatal Straight Flow Sensor ISO15 8411130 For Straight Flow Sensor Babylog 8000 Evita4 Evita2 Dura NeoFlow

    Ask Lasest Price
    Brand Name : Dräger
    Model Number : 8411130
    Payment Terms : T/T, Western Union,PayPal
    Delivery Time : 3-5 working days
    • Product Details
    • Company Profile

    Drager Neonatal Straight Flow Sensor ISO15 8411130 For Straight Flow Sensor Babylog 8000 Evita4 Evita2 Dura NeoFlow


    Drager Neonatal Straight Flow Sensor ISO15 8411130 for Straight flow sensor for Babylog® 8000/Evita® 4/ Evita 2 dura®/Evita® NeoFlow.


    ►Brief Introduction

    The measurement principle used in flow measurement is the hot wire velocimetry method.Hot wire velocimetry is a thermal measurement method in which the measuring line of the flow meter remains at a constant high temperature during the measurement process. The higher the flow rate, the more energy is required to maintain a constant high temperature. The flow rate can then be calculated from the energy consumed. To ensure proper functionality, check for visible damage, dirt and particles before inserting the flow meter. Repeat this check regularly. If the flow meter is damaged, dirty, or contains particles, replace it. If the measurement line of the flow meter continues to glow during operation, it is a sign of dirt. Reprocess or replace the flow meter immediately.


    Work for Draeger Inc.

    • BABYLOG 8000
    • EVITA 4
    • EVITA XL
    • BABYLOG
    • EVITA 4/2D

    ► Safety-related information

    Flow sensors

    Flammable substances

    The flow sensor may ignite medications or other substances based on easily flammable substances. The patient may be put at risk.

    • Do not nebulize medications or other substances that are easily flammable or spray them into thedevice.
    • Do not use substances containing alcohol.
    • Do not allow combustible or explosive substances to enter the breathing systemor the breathingcircuit.

    Residual vapors of highly flammable disinfectants

    Residual vapors of highly flammable disinfectants (e.g., alcohols) and deposits that were not removed during reprocessing may ignite when the flow sensor is in use.

    The patient may be put at risk.

    • Ensure particle-free cleaning and disinfection.
    • After disinfection, allow the flow sensor to air-dry for at least 30 minutes.
    • Before inserting, check the flow sensor for visible damage and soiling, such as residual mucus,medication aerosols, and particles.
    • Replace flow sensors when damaged, soiled, or not particle-free.Neonatal flow sensor improper reprocessing and soiling, such as deposits or particles, may damage the flow sensor. The flow measurement may fail. As a result, the patient may be put at risk.
    • No machine cleaning or disinfection of the sensor insert
    • No plasma sterilization or radiation sterilization
    • No water jets, compressed air, brushes or the like when cleaning the sensor insert
    • No ultrasonic bath

    ► Steps

    1. Pull out the flow meter (6) from the Y-piece (5).

    2. Unplug the flow meter cable plug (7) from the flow meter (6).

    3. While gently pressing the knobs (8) on both sides, pull the flow meter insert (9) out of the flow meter housing (10).


    Specification

    ItemNEONATAL FLOW SENSOR ISO15
    OEMDraeger Inc.
    OEM #8411130
    ReturnableNo
    StandardsISO 15

    If you have any questions about Drager Neonatal Straight Flow Sensor ISO15 8411130 for Straight flow sensor for Babylog® 8000/Evita® 4/ Evita 2 dura®/Evita® NeoFlow., please feel free to contact us at any time, we are glad to answer them. Looking forward to receiving your early inquiry. Thanks for Your Cooperation.


    Quality Drager Neonatal Straight Flow Sensor ISO15 8411130 For Straight Flow Sensor Babylog 8000 Evita4 Evita2 Dura NeoFlow for sale
    • Haven't found right suppliers
    • Our buyer assistants can help you find the most suitable, 100% reliable suppliers from China.
    • And this service is free of charge.
    • we have buyer assistants who speak English, French, Spanish......and we are ready to help you anytime!
    Submit Buying Request
    Send your message to this supplier
    *From:
    Your email address is incorrect!
    *Subject:
    Your subject must be between 10-255 characters!
    *Message:
    For the best results, we recommend including the following details:
    • --Self introduction
    • --Required specifications
    • --Inquire about price/MOQ
    Your message must be between 20-3,000
    Yes! I would like your verified suppliers matching service!
    Send your message to this supplier
     
    *From:
    *To: China World Technology Medical Equipment Service Group
    *Subject:
    *Message:
    Characters Remaining: (0/3000)
     
    Explore more Patient Monitor Accessories products from this supplier
    Inquiry Cart 0